CDSCO Medical Device Regulations and Registration Process in India

Medical Device Regulations and Registration Process in India

Planning to import or manufacture a medical device in India? You need approval first. CDSCO medical device registration is that approval, whether it comes as a licence, a permission or a registration under the Medical Devices Rules, 2017. Which one applies to you depends on the device’s risk class and its intended purpose, and also on whether you import it or manufacture it.

CDSCO is the central regulator. Still, state drug controllers take care of part of the work, so not every application ends up at CDSCO headquarters. That’s why knowing the right authority early matters.

This guide covers who needs which licence, the application process, documents, fees, timelines, technical requirements and post-approval duties.

CDSCO and Its Role in Medical Device Regulation

CDSCO is the national regulator under the Ministry of Health and Family Welfare. It does not process every application. The authority depends on the device class and the activity.

  • Central Licensing Authority (CLA): It is the Drugs Controller General of India. It handles import of all classes, manufacture of Class C and D, clinical investigations and clinical performance evaluation of new IVDs.
  • State Licensing Authorities (SLAs): They are the State Drugs Controllers. They handle manufacture of Class A and B devices and the sale, stock or distribution of all classes.

In practice, an importer deals with CDSCO. A Class B manufacturer applies to its state authority. A Class D manufacturer applies to CDSCO. Filing with the wrong authority costs weeks.

The licence holder is responsible for quality, labelling, adverse event reporting and recalls.

Medical Device Regulations in India

Medical devices are regulated as drugs under the Drugs and Cosmetics Act, 1940. The detailed rules are in the Medical Devices Rules, 2017 (MDR 2017), issued as G.S.R. 78(E) and in force from 1 January 2018.

CDSCO defines a medical device broadly. Any instrument, apparatus, implant, software or accessory intended for the diagnosis, prevention, monitoring, treatment or alleviation of disease or injury qualifies. So do devices that support or sustain life, disinfecting devices and devices that control conception. One group falls outside it. Products that achieve their primary action by pharmacological, immunological or metabolic means are not medical devices.

All devices fall within the regulated definition. S.O. 648(E) of 11 February 2020 notified every product meeting the definition of a medical device as a drug under the Drugs and Cosmetics Act, 1940.

Licensing became mandatory in two phases. Class A and B devices came first, from 1 October 2022. Class C and D followed from 1 October 2023. All four risk classes now fall under the applicable licensing requirements.

The Rules have changed several times since 2017, and one amendment matters for Class A. G.S.R. 777(E) of 14 October 2022 took non-sterile and non-measuring Class A devices out of the licensing regime and placed them under online registration. CDSCO publishes the Rules and their amending notifications on its Medical Devices Rules page.

Medical Device Classification in India

Medical devices are classified into four risk-based categories (Class A, B, C and D). Classification depends on the device’s intended purpose, duration of use, degree of invasiveness and potential risk to patients, rather than its product name.

Class Risk level
A Low
B Low-moderate
C Moderate-high
D High

The Central Licensing Authority (CLA) classifies devices and publishes class-wise lists on the CDSCO website, organised by specialty. It can add, remove or change entries at any time. Class A non-sterile and non-measuring devices form a separate group. They are registered online rather than licensed.

Class matters more than anything else here. It sets the licence, the authority, the audit route and the document set.

CDSCO Medical Device Registration and Licensing Requirements

“CDSCO medical device registration” is the phrase most people search, but for most devices the legal instrument is a licence. A licence is granted in Form MD-5, MD-6, MD-9, MD-10 or MD-15. Registration applies only in specific cases.

Import licence (MD-15)

The CLA grants it for all classes. The application is made by an Indian authorised agent that holds a manufacturing licence, a wholesale licence or an MD-42 registration certificate. A foreign manufacturer does not apply directly.

Manufacturing licence

The licensing authority depends on the device’s risk classification:

  • Class A (other than non-sterile and non-measuring) and Class B: Apply to the State Licensing Authority (SLA).
  • Class C and D: Apply to the Central Licensing Authority (CLA).

Applicants must have a manufacturing site meeting the applicable quality management requirements and employ qualified technical staff.

Loan Licence

A loan licence allows a manufacturer to manufacture a medical device using the manufacturing facilities of another licensed manufacturer. It follows the same licensing authority structure as regular manufacturing licences.

Class A non-sterile and non-measuring devices

These are not licensed. The manufacturer or importer uploads device details, an undertaking on classification, self-certification to the essential principles and standards, and a signed declaration on the Online System for Medical Devices. Importers also upload an overseas site registration or Free Sale Certificate. A registration number is generated. Authorities can still verify records and suspend or cancel the registration. Other Class A devices, such as sterile or measuring ones, follow the licence route.

Special permissions

The MDR 2017 provides specific licences and permissions for the following purposes:

  • Test licences for clinical investigation, evaluation or training: MD-12/MD-13 (manufacture) and MD-16/MD-17 (import).
  • Devices with no predicate: MD-26 and MD-27 (Rule 63).
  • Investigational devices imported by government hospitals for patient treatment: MD-18/MD-19.

Responsibilities of Foreign Manufacturers and Indian Agents

Two parties share the work on imported devices. The foreign manufacturer supplies the product and manufacturing site documentation. Its Indian authorised agent applies for the import licence and holds it. Importers and distributors must also follow the applicable labelling, record-keeping and sale requirements.

CDSCO Medical Device Registration Process in India: Step by Step

The process involves the following steps, from determining regulatory status and risk classification to obtaining the applicable licence or registration and meeting post-approval obligations.

Medical Device Registration Steps In India

Step 1: Identify whether the product is a medical device

Start by checking the product against the CDSCO definition. Is it a device, a drug, an IVD or another regulated product? CDSCO lists clarification on regulatory status among its functions. Never decide status without full product information.

Step 2: Determine the risk class

Use the CDSCO class lists. A Class A non-sterile and non-measuring device follows online registration, not the licence steps below.

Step 3: Identify the application and licence forms

Route Application form Licence form Authority
Import MD-14 MD-15 CLA
Class A/B manufacturing MD-3 MD-5 SLA
Class A/B loan licence MD-4 MD-6 SLA
Class C/D manufacturing MD-7 MD-9 CLA
Class C/D loan licence MD-8 MD-10 CLA

Step 4: Appoint an Indian Authorised Agent (import only)

An authorised agent is appointed by the overseas manufacturer through a power of attorney to import into India. The agent files the MD-14 application, pays fees and carries licence conditions. The licensee must inform CDSCO within 15 days of any foreign regulatory action such as a market withdrawal, and must report changes in the constitution of the firm within 30 days. Domestic manufacturers do not need an agent.

Step 5: Prepare the documentation

The set depends on application type, class, product type and whether you import or manufacture.

Step 6: Register on the CDSCO online system

CDSCO’s pages refer to the Sugam online portal. CDSCO’s regulatory chart names the Medical Devices online portal (cdscomdonline.gov.in) and the National Single Window System (nsws.gov.in). Confirm the correct entry point for your application type before filing.

Step 7: Complete and submit the application

Complete the form for your route, upload documents in checklist order, make sure device names, models, sites and class match across every file, then submit and track the application online.

Step 8: Pay the government fees

Fees depend on the application type, class, number of sites and number of devices.

Step 9: Respond to regulatory queries

The licensing authority may come back with questions. They can ask you to clarify a point or to send additional documents, and a quick, complete reply keeps the file moving. Class C and D applications carry more risk here. If the documents are incomplete, the CLA can reject the application outright. When the deficiencies can be rectified, the review clock restarts on the date the applicant removes them.

Step 10: Complete inspection or audit, if required

  • Class B sites are audited by a registered Notified Body before the licence is granted.
  • Class C and D sites are inspected by the licensing authority or a Notified Body.
  • For import, the CLA may inspect the overseas site.

Not every applicant is inspected.

Step 11: Obtain the licence or permission

The outcome is a licence (MD-5, MD-6, MD-9, MD-10 or MD-15), a permission (MD-27), or, for Class A non-sterile and non-measuring devices, a registration number. Each carries its own conditions.

Step 12: Meet post-approval requirements

Compliance continues after the licence is granted.

Documents Required for CDSCO Medical Device Registration

The exact documents depend on the application type, device class, product category and the applicable CDSCO checklist.

Documents for an import licence

The CDSCO checklist for MD-14 lists:

  • Covering letter, application form and fee challan.
  • Power of attorney with the agent’s undertaking, authenticated by a first-class Magistrate, the Indian Embassy or by apostille.
  • Copy of the agent’s wholesale or manufacturing licence or MD-42 certificate, plus constitution details.
  • Regulatory certificates: Free Sale Certificate or marketing authorisation, overseas site registration, and the latest inspection report from the competent authority within three years, if any.
  • Quality certificates for the actual manufacturing site and the manufacturer’s declaration of conformity.
  • Plant Master File.
  • Device Master File: device description, grouping justification, specifications, labelling and instructions for use, Essential Principles checklist, risk analysis summary, verification and validation, and, where applicable, biocompatibility, sterilisation, software, stability and clinical data, post-marketing data, and batch release certificates for three consecutive batches.

Documents for a manufacturing licence

The following documents must be prepared and submitted as part of the manufacturing licence application:

  • manufacturing site details
  • Device Master File and Plant Master File
  • quality management documents
  • risk management records
  • testing and validation reports
  • Essential Principles compliance records
  • product labels

CDSCO Medical Device Registration Fees

The applicable government fees are prescribed under the Second Schedule of the MDR 2017.

The fee structure is summarised below.

Application Class Per site Per device
Manufacturing licence A (other than non-sterile and non-measuring) or B ₹5,000 ₹500
Manufacturing licence C or D ₹50,000 ₹1,000
Import licence, non-IVD A (other than non-sterile and non-measuring) $1,000 $50
Import licence, non-IVD B $2,000 $1,000
Import licence, non-IVD C or D $3,000 $1,500
Import licence, IVD A (other than non-sterile and non-measuring) or B $1,000 $10
Import licence, IVD C or D $3,000 $500

An overseas site inspection carries a separate fee of $6,000. Retention fees follow the same site-and-device structure. Check the current sheet before paying, because fees can change by amendment.

CDSCO Medical Device Approval Timelines

The MDR 2017 prescribes timelines for regulatory decisions but does not specify a single overall timeline for completing the entire approval process.

Pathway Stage and timeline
Class A manufacturing (other than non-sterile and non-measuring) Decision within 45 days of application
Class B manufacturing
  • Audit within 90 days
  • Audit report within 30 days
  • Decision within 20 days of the report
Class C and D manufacturing
  • Scrutiny within 45 days
  • Inspection within 60 days of application
  • Decision within 45 days of the inspection report
Import licence Grant or rejection within 9 months of application
Devices with no predicate 120 days, extendable by up to 30 days

The time required for application preparation and responding to regulatory queries is not included in these periods.

Technical and Clinical Requirements for Medical Devices

The following requirements and considerations apply across the relevant regulatory pathways, depending on the device type and intended purpose.

  • Essential Principles: Manufacturers must comply with the Essential Principles of Safety and Performance specified in MoHFW guidelines. CDSCO publishes guidance on these principles, and the Device Master File includes an Essential Principles checklist.
  • Quality management system: Manufacturers must establish and maintain a quality management system. For imports, CDSCO’s checklist requires a quality management system certificate or applicable CE quality assurance certificates. ISO 13485 is a recognised standard for establishing and demonstrating compliance with quality management requirements.
  • Standards: Devices must conform to applicable BIS or MoHFW-notified standards. If none exist, ISO or IEC standards or pharmacopoeial standards apply. After that, validated manufacturer’s standards apply.
  • ISO 14971: A risk analysis and control summary must be included in the documentation. ISO 14971 provides a recognised framework for medical device risk management.
  • Testing and validation: The Device Master File covers verification and validation and, where applicable, biocompatibility, sterilisation, software and stability data. The CLA can order testing at the applicant’s cost if quality is in doubt.
  • Clinical evidence for import: With a Free Sale Certificate from Australia, Canada, Japan, EU countries, the UK or the USA, a licence can be granted without clinical investigation in India. For other countries, Class C and D devices may need a clinical investigation in India. For Class A (other than non-sterile and non-measuring) and Class B, published safety and performance data, or a clinical investigation in the country of origin, with a Free Sale Certificate, may be accepted.
  • Devices without a predicate: a predicate is a first-of-its-kind device already approved by the CLA. Without one, manufacturers must obtain the applicable regulatory permission following the required clinical investigation. Certain exemptions may apply to devices approved and marketed for at least two years in listed countries.
  • IVDs: New IVDs need permission for clinical performance evaluation (MD-24/MD-25), then permission to import or manufacture (MD-28/MD-29). See the CDSCO IVD page for lists approved performance evaluation laboratories.
  • Software and AI: CDSCO released a draft Guidance Document on Medical Device Software in October 2025, covering software in a medical device (SiMD) and software as a medical device (SaMD). It states that software is regulated and classified according to its intended medical purpose under the existing MDR 2017 framework. No separate AI approval route is currently prescribed.

Post-Approval Compliance for Medical Devices in India

A licence stays valid only while you meet its conditions and pay retention fees. Licence holders must continue to meet the following post-approval requirements:

  • Labelling and packaging: Mandatory label content includes device name, manufacturer and premises, net quantity, manufacture and expiry dates, batch number, storage conditions and sterility. Domestic devices carry the manufacturing licence number. Imported devices carry the import licence number, importer details, actual manufacturing premises and manufacture date, by sticker if not printed. Imports must also meet minimum residual shelf life.
  • Import and distribution: Each import consignment must travel with an invoice or statement showing device name and quantity. Devices must be supplied with a package insert or user manual, where applicable.
  • Manufacturing and quality records: Licence holders must test each batch before release, maintain an audit book in Form MD-11, retain reference samples of invasive devices and IVDs for 180 days after expiry, and keep manufacturing and sales records available for inspection.
  • Traceability: Unique Device Identification (UDI) requirements apply from the date notified by the Central Government.
  • Post-market surveillance and materiovigilance: Post-market surveillance involves the systematic collection and analysis of information on marketed medical devices. Licence holders must report suspected unexpected serious adverse events and recalls within 15 days of becoming aware of them. Reports are submitted using the Medical Device Adverse Event Reporting Form (Version 1.2) under the Materiovigilance Programme of India. CDSCO also provides a Field Safety Corrective Action form on its Medical Device & Diagnostics page.
  • Recalls and field actions: the licensing authority can direct withdrawal or recall of batches.
  • Changes: Changes to approved medical devices are classified as major or minor. Major changes require prior approval, while minor changes must be reported within 30 days. The authority must respond within 45 days for manufacturing licences and 60 days for import licences, failing which approval is deemed granted. Changes in the constitution of a licence holder require a fresh application within 180 days. Not every change requires a new licence.
  • Retention: Licences stay valid in perpetuity if the retention fee is paid before five years from issue. Late payment adds 2 per cent per month, for up to 180 days on manufacturing licences and 90 days on import licences. After that the licence is deemed cancelled.

Post-approval compliance also requires manufacturers to stay informed about changes in CDSCO regulations and guidelines. Artixio’s Qurious RI helps manufacturers track regulatory updates from health authorities worldwide and receive personalized alerts on relevant changes, making it easier to stay informed and maintain ongoing compliance.

Conclusion

CDSCO medical device registration starts with the Medical Devices Rules, 2017 and your device’s class. The class decides the authority, the form and the documents. Choosing the right route matters just as much: MD-14 for import, MD-3 or MD-7 for manufacturing, or online registration for Class A non-sterile and non-measuring devices. A complete Device Master File and Plant Master File prevent most query rounds.

A licence is only one part of compliance. Retention fees, change approvals, labelling and adverse event reporting continue for as long as you sell in India. To identify the right regulatory pathway for your device, explore Artixio’s medical device regulatory services in India or contact us at info@artixio.com.

Frequently Asked Questions

1. What is CDSCO medical device registration?

It refers to obtaining the applicable licence, permission or registration under the Medical Devices Rules, 2017. Most medical devices require a licence, while Class A non-sterile and non-measuring devices follow an online registration process.

2. What are the four medical device classifications in India?

Medical devices are classified into four risk categories under Rule 4 of the MDR 2017: Class A (low risk), Class B (low-moderate risk), Class C (moderate-high risk) and Class D (high risk).

3. Which forms are required for medical device manufacturing and import?

For manufacturing, Class A and B applications use MD-3 (MD-4 for loan licences), while Class C and D applications use MD-7 (MD-8 for loan licences). Import applications use MD-14, with the licence issued in MD-15.

4. How much does CDSCO medical device registration cost?

Fees depend on the licence type, risk classification, number of manufacturing sites and number of devices. For example, a Class B non-IVD import licence costs USD 2,000 per site plus USD 1,000 per device.

5. How long does the CDSCO medical device approval process take?

The MDR 2017 prescribes statutory timelines for specific regulatory decisions. For example, the import licence decision period is nine months. The overall approval timeline may vary depending on application preparation and regulatory queries.