IVD Regulation and Registration Process in Mexico

IVD Regulation and Registration Process in Mexico

IVD classification and registration in Mexico starts with COFEPRIS classifying in vitro diagnostics as medical devices, which require sanitary registration before production, sale or distribution unless included on the low-risk list.The risk class (I, II or III) sets the evidence COFEPRIS expects, the fee and the response period, so a wrong class affects every later step.

This article explains COFEPRIS IVD registration for manufacturers, including foreign companies that act through a Mexican applicant. It covers classification, the application and technical dossier, labeling, import and establishment duties and post-market obligations.

Regulatory Framework for IVDs in Mexico

Role of COFEPRIS in IVD Regulation

COFEPRIS is the Ministry of Health’s regulator for medical devices, IVDs included. COFEPRIS evaluates, issues and revokes sanitary registrations. It grants one only when the applicant has met the applicable requirements and paid the fees. It supervises the import and export of these products. It also checks labels at manufacturing sites, warehouses and distributors, and during review of registration applications and their modifications.

Key Mexican Regulations Governing IVD Products

  • Ley General de Salud (LGS): Sets the legal basis for medical devices and related regulatory requirements.
  • Reglamento de Insumos para la Salud (RIS): Covers sanitary registration requirements and the three medical device risk classes.
  • FEUM Medical Devices Supplement 5.0: Provides classification rules and registration guidelines.
  • Mexican Official Standards (NOMs): Set requirements for labeling, good manufacturing practices and technovigilance.
  • COFEPRIS Agreements and Guides: Explain the technical documentation and procedures required for sanitary registration.

Regulatory Definition and Scope of IVD Products

NOM-137-SSA1-2025 provides the official definitions of diagnostic agents and in vitro diagnostic devices:

  • Agente de diagnóstico: Includes antigens, antibodies, calibrators, controls, reagents, reagent kits, culture media and similar products used in clinical or paraclinical procedures.
  • IVD (agente de diagnóstico in vitro): A medical device intended for the in vitro examination of human samples to provide information for diagnosis, monitoring or compatibility. This includes reagents, calibrators, sample collection devices, control materials and related instruments.
  • Self-testing IVDs: Products intended for use by individuals without formal medical or laboratory training.
  • Rapid tests: Screening tests that may require confirmation through laboratory or clinical testing.

Research-use-only and general laboratory products: Not all laboratory products qualify as medical devices. Their regulatory status depends on their intended purpose and use. Examples of products that are not considered medical devices and do not require sanitary registration include laboratory buffers, chemical reagents, staining kits for histology and culture media for industrial or food microbiology.

Manufacturers should assess the intended use of each product carefully, as a research-use-only label alone does not automatically exempt it from COFEPRIS requirements.

Mexico’s IVD regulations continue to evolve, making it important for manufacturers to stay informed about COFEPRIS updates. Artixio’s QuriousRI regulatory intelligence platform helps regulatory teams track changing requirements through personalized alerts, expert-reviewed summaries, and risk-impact analysis. Explore the platform to learn how it can support your regulatory compliance activities.

IVD Risk Classification in Mexico

IVD classification in Mexico is based on the product’s intended purpose and associated risk.

Class General definition (RIS Art. 83) IVD examples in FEUM Appendix II
I Known in medical practice, safety and efficacy proven, generally not introduced into the body
Reagents, calibrators, controls and culture media for a specific test; lay-user tests whose result does not determine a critical situation
II Known, may vary in materials or concentration, generally introduced for under 30 days
Blood group and tissue typing for compatibility; sexually transmitted infections, certain cancer, genetic, cardiac, prenatal and newborn screening tests; SARS-CoV-2; lay-user and rapid tests
III New or recently accepted, or introduced and remaining over 30 days
HIV, HCV, HBV and HTLV tests; screening for transfusion or transplant suitability; ABO, Rh, Kell, Kidd and Duffy typing

Mexico follows its own regulatory criteria for IVD classification, and manufacturers must meet COFEPRIS requirements before submitting an application. Class I products are not automatically exempt from sanitary registration; only products included in the official low-risk list qualify for exemption.

Factors Affecting IVD Classification

IVD classification in Mexico depends primarily on the manufacturer’s declared intended purpose. When multiple classification rules apply, the rule resulting in the higher risk class takes precedence. Devices used in combination are generally classified separately.

The main factors affecting IVD classification include:

  • Intended purpose: Whether the product is used for screening, diagnosis, monitoring or compatibility testing.
  • Target analyte or condition: The specific disease, infection or biological marker being detected.
  • Specimen type: Whether the test uses blood, cerebrospinal fluid, tissues or other human samples.
  • Potential consequences of incorrect results: The risk of death, serious disability or disease transmission.
  • Intended user: Whether the product is designed for healthcare professionals or lay users.
  • Testing environment: Whether the IVD is intended for laboratory or near-patient testing.
  • Related products: Calibrators and controls may follow the classification of their associated reagents, while a kit may take the classification of its highest-risk component.

Manufacturers should establish the intended purpose before classification, as changes to the intended use can affect the risk class, supporting evidence and labeling requirements.

COFEPRIS IVD Registration and Authorization Requirements

Before commercialization in Mexico, an IVD generally requires a sanitary registration unless it is included on the official list of low-risk devices exempt from registration or is not classified as a medical device.

Sanitary Registration Requirements

Sanitary registration authorizes the medical device for marketing in Mexico. COFEPRIS processes applications under the procedure Registro Sanitario de Dispositivos Médicos (homoclave COFEPRIS-04-060), which applies to both domestically manufactured and imported products. Low-risk devices that still require registration may qualify for the simplified procedure COFEPRIS-04-070

  • Application form and fees: Completed Form FF-COFEPRIS-21 and proof of payment. Fees vary according to the device classification and the applicable provisions of the Ley Federal de Derechos.
  • Labeling and instructions: A draft label in Spanish that complies with the applicable NOM, along with Spanish instructions for use where required.
  • Technical documentation: Scientific and technical information demonstrating the product’s safety, efficacy and quality, as specified in the COFEPRIS registration guide.
  • Foreign-manufactured devices: A Certificate of Free Sale or equivalent document issued by the authority in the country of origin, a GMP certificate or equivalent document, and a manufacturer representation letter when the foreign manufacturer is not the applicant’s parent company or affiliate. For electronically issued certificates, the application may include the relevant access route.

The COFEPRIS Guide provides separate documentation checklists for Class I, II and III devices. The IVD’s risk class determines the technical documentation required. Applicants must also submit a power of attorney, where applicable.

Low-Risk IVD Products and Possible Exemptions

The Acuerdo published on July 7, 2025 replaced the previous 2011 and 2014 lists and introduced three categories of products:

  • Annex One (93 devices): Covers low-risk devices that still require sanitary registration but benefit from simplified requirements and a resolution period of 20 business days.
  • Annex Two (104 devices): Covers low-risk devices that do not require sanitary registration. These products must still comply with good manufacturing practices and technovigilance requirements. Importers must declare the product’s eligibility to customs under oath. They must also have an Notice of Operation (Aviso de funcionamiento) and a Sanitary Responsible Person (Responsable sanitario).
  • Annex Three: Lists products that are not considered medical devices and therefore do not require sanitary registration.

For IVD manufacturers, being classified as Class I does not automatically mean exemption from registration. Low-risk classification is subject to specific conditions. The official low-risk lists mainly cover general medical equipment and supplies, such as dressings, beds, centrifuges and microscopes, rather than IVD reagent kits. Manufacturers should check the applicable lists against their specific products before determining whether sanitary registration is required.

Registration Compared with Other Authorizations

Authorization What it covers Key details
Sanitary registration The product
COFEPRIS-04-060; valid five years, extendable
Aviso de funcionamiento or sanitary license, plus responsable sanitario The company and its premises
Applies to importers and distributors according to their activity
Import permit The shipment
“Permiso sanitario de importación de dispositivos médicos” with or without registration (COFEPRIS-01-014 and 01-015)

Registration authorizes the product. It does not license the company or replace the permit for a shipment.

Also Read: Medical Device Regulations and Registration Process in Mexico

Requirements for Foreign IVD Manufacturers Entering Mexico

Foreign manufacturers looking to enter the Mexican market generally need a local applicant and legal representative to handle the sanitary registration process. A power of attorney must be provided for the legal representative established in Mexico.

The main requirements include:

  • Mexican applicant and legal representative: A Mexican representative submits and signs the registration application on behalf of the manufacturer. The required legal authorization must also be provided.
  • Registration holder: The registration holder (titular del registro sanitario) is the person or company that obtains the sanitary authorization. Foreign manufacturers should confirm the applicable requirements with COFEPRIS before deciding how to structure the application.
  • Manufacturer documents: The foreign manufacturer must provide a Certificate of Free Sale, GMP evidence and, where applicable, a manufacturer representation letter.
  • GMP evidence: COFEPRIS accepts GMP certificates issued by the authority in the country of origin or equivalent documents. ISO 13485 and MDSAP documentation may also be accepted under applicable Mexican requirements.
  • Importer and distributor: Importers and distributors have separate establishment obligations. They must also meet the applicable labeling requirements.
  • Coordination between both parties: The foreign manufacturer is responsible for providing the technical documentation, while the Mexican applicant manages the local registration process and responds to COFEPRIS queries. It is best to agree on document ownership, translations and label versions before submitting the application.

Step-by-Step IVD Registration Process in Mexico

Step 1. Define the intended purpose

The analyte, specimen, target population, user, setting and every diagnostic claim should be written down. Classification, evidence and labeling all follow from this text.

Step 2. Determine the risk class

The FEUM IVD classification rules are applied to the stated purpose. Where more than one rule fits, the rule giving the higher class applies. Calibrators, controls and kits are classified as described above.

Step 3. Confirm the procedure

First, determine whether the IVD requires sanitary registration. If registration is required, the applicable procedure is COFEPRIS-04-060 or COFEPRIS-04-070 for eligible low-risk devices. COFEPRIS also offers an abbreviated, reliance-based route that considers decisions from reference regulatory authorities. Manufacturers should confirm whether this option applies to their IVD. The August 2025 Acuerdo replaced the earlier separate equivalence procedures for the FDA, Health Canada and Japan.

Step 4. Identify the Mexican applicant

The company that will hold the registration is chosen and its legal representative is appointed.

Step 5. Prepare the technical and administrative file

The COFEPRIS Guide has a separate checklist for Class I, II and III devices, so the device class decides how much evidence the file must contain. The file combines administrative documents, such as the application form, proof of payment and legal representation, with technical evidence on specifications, stability, performance and quality. Each item should be matched against the checklist for the declared class before submission, since missing documents are a common reason COFEPRIS issues information requests.

Step 6. Prepare Spanish labeling and instructions for use

The draft label must be in Spanish and carry the product names, expiry date, lot number, storage conditions, warnings and the “agente de diagnóstico in vitro” legend. Imported products also show the importer’s details and country of origin. The 2008 labeling standard applies until May 14, 2027, and products marketed after that date need labels that can meet the 2025 version.

Step 7. Submit to COFEPRIS

The application is filed through the channel shown on the COFEPRIS-04-060 procedure page, with form FF-COFEPRIS-21 and the documents.

Step 8. Pay fees and answer queries

The fee for the device class is paid under the Ley Federal de Derechos. COFEPRIS can ask for missing information. For low-risk devices that still require registration, the Acuerdo sets 10 business days for administrative requests and 15 for technical ones, and gives the applicant 10 business days to answer.

The August 22, 2025 Acuerdo lowers COFEPRIS’s maximum response periods to 20 business days (Class I), 25 (Class II) and 35 (Class III), from 30, 35 and 60, and to 15 for low-risk devices. These figures are administrative maximums. They do not estimate actual duration. The reduction applies 180 business days after publication, and only once the needed digital tools are running. Until COFEPRIS confirms the tools are operating, the earlier periods should be used for planning.

Step 9. Obtain and verify the authorization

The registration must be checked against COFEPRIS’s official device registry listings (COFEPRIS). NOM-137 requires the registration number on the label or counter-label.

Step 10. Maintain compliance.

Marketed labels, specifications and manufacturers must stay identical to what COFEPRIS approved, and changes are filed before they are applied.

Technical Documentation Required for IVD Registration in Mexico

The technical dossier for IVD registration in Mexico depends on the product’s risk class, intended purpose and registration procedure. Manufacturers should prepare the documents relevant to their product and check the applicable COFEPRIS requirements before submitting the application.

The application generally includes:

  • Product description and intended purpose, including components, models and presentations.
  • Reagent composition and characteristics, where applicable.
  • Product specifications and supporting test results.
  • Stability studies demonstrating product shelf life, particularly for diagnostic reagents.
  • Analytical and, where applicable, clinical performance evidence.
  • Manufacturing and quality information, including GMP certification or equivalent documentation.
  • Draft labeling and instructions for use in Spanish.
  • Certificate of Free Sale and manufacturer representation letter for foreign-made products, where required.
  • Proof of legal representation and payment of applicable fees.

Manufacturers must also maintain certain technical and quality records and make them available to COFEPRIS when required. These include:

  • Risk management records covering identified risks, evaluations and residual risks.
  • Document-control records for labeling versions and changes.
  • Verification and validation records, including supporting performance study documentation.
  • Technovigilance and complaint records.

These records should remain updated and accessible. Not every supporting document needs to be submitted with the initial application. The filing requirements depend on the IVD’s risk class and applicable COFEPRIS requirements.

Analytical and Clinical Performance Requirements for IVD Products

Performance evidence is an important part of IVD registration in Mexico. The information provided must support the claims made on the product label and demonstrate that the test performs as intended. NOM-137 distinguishes between analytical and clinical performance.

Analytical Performance

Analytical performance refers to how accurately an IVD detects or measures a specific analyte. NOM-137 requires manufacturers to include relevant performance characteristics in the instructions for use (Section 5.10.12.11), including:

  • Precision and accuracy.
  • Analytical sensitivity and specificity.
  • Potential interfering substances, such as hemolysis and lipemia.
  • Traceability of calibrator values to reference materials or procedures.
  • Stability during use, including storage conditions for working solutions.
  • Quality-control procedures and criteria for evaluating results.

Other characteristics, such as detection limits, measuring range and cross-reactivity, may also need to be addressed depending on the product and its intended claims.

Clinical Performance

Clinical performance demonstrates how well an IVD’s results relate to a clinical condition in the intended population. Where applicable, NOM-137 requires supporting clinical performance information in the instructions for use (Section 5.10.12.12). This may include:

  • Diagnostic sensitivity and specificity.
  • Positive and negative predictive values.
  • Likelihood ratios.
  • Expected values in healthy and affected populations.

The type and extent of clinical evidence depend on the IVD, its intended purpose and risk class, as well as the applicable COFEPRIS requirements. Manufacturers should ensure that their performance studies adequately support the diagnostic claims made for the product.

IVD Labeling and Instructions for Use Requirements in Mexico

IVD labels and instructions for use must be in Spanish, comply with NOM-137 and match the information authorized by COFEPRIS. Manufacturers should ensure that product labeling meets Mexican requirements before commercialization.

NOM-137-SSA1-2025 was published in the DOF on May 19, 2026. It will take effect on May 14, 2027, replacing NOM-137-SSA1-2008. A 180-day transition period applies to certain existing labeling and finished stock.

The main labeling requirements include:

  • Generic and brand names, along with the COFEPRIS registration number.
  • Manufacturer details and, for imported products, importer information and country of origin.
  • Manufacturing date, expiry date and lot or serial number.
  • Catalogue, model or product reference.
  • Package contents, such as the number of tests or reagent volume.
  • Storage conditions, warnings and precautions.
  • The required IVD designation and applicable symbols.

Instructions for use must provide relevant information about the test, including its intended purpose, specimen requirements, testing procedure, result interpretation, quality control and analytical and clinical performance. Self-testing products must also include clear instructions for users and explain when medical advice is needed.

Foreign labels accepted in the US or EU do not automatically meet Mexican requirements. Imported IVDs must comply with Spanish-language labeling and applicable Mexican registration and importer information requirements. Missing information may be added through a counter-label in Mexico, where permitted by NOM-137.

IVD Import Requirements and Commercialization in Mexico

Importing and commercializing IVD products in Mexico involves more than obtaining sanitary registration. Manufacturers and importers must also meet applicable import, establishment, storage and labeling requirements.

The main requirements include:

  • Sanitary registration: IVDs generally require sanitary registration before sale or distribution, unless they qualify for an applicable exemption.
  • Import permit: COFEPRIS issues sanitary import permits for medical devices, with or without registration, depending on the applicable procedure. The August 11, 2025 Acuerdo simplified these procedures. Products exempt from registration may also be exempt from import permits, subject to the applicable conditions.
  • Importer requirements: Importers must have the required Notice of Operation (Aviso de funcionamiento) or sanitary licence and a Sanitary Responsible Person (Responsable sanitario). Their name and address must also appear on the product label.
  • Manufacturing and storage: NOM-241-SSA1-2025 establishes Good Manufacturing Practice requirements for establishments involved in manufacturing, storing and distributing medical devices in Mexico. It has been in force since November 30, 2025.
  • Pre-shipment checks: Before shipping, manufacturers and importers should verify the product’s registration status and ensure that the labeling matches the COFEPRIS authorization.
  • Responsibilities: The registration holder, importer and distributor may be different companies. Their respective regulatory responsibilities should be clearly defined.

Sanitary registration and import authorization are separate procedures. Obtaining registration does not automatically authorize every shipment entering Mexico.

Post-Market Compliance and Changes to IVD Registrations

Obtaining sanitary registration is not the end of the regulatory process. IVD manufacturers and registration holders must continue meeting post-market requirements and keep their registration details up to date.

Technovigilance and complaints

Technovigilance is mandatory for registered medical devices in Mexico. Manufacturers and registration holders must monitor product-related incidents and comply with the reporting requirements under NOM-240-SSA1-2012. NOM-137 also requires labeling to include contact information for reporting incidents.

Registration validity and renewal

Sanitary registration is valid for five years and can be renewed for additional five-year periods. Renewal requires the applicable application form, proof of fee payment and supporting documents. Foreign manufacturers may also need to provide a valid ISO 13485 certificate or equivalent GMP documentation and a manufacturer representation letter, where applicable.

Changes to registered products

Any changes to an approved IVD registration must follow the applicable COFEPRIS modification procedure. The requirements depend on the type of change:

  • Administrative modifications: Include changes to the registration holder’s address, distributor details and certain packaging information.
  • Technical modifications: Apply when changes affect the product’s technical characteristics, specifications or other approved conditions.
  • Immediate-resolution modifications: Certain administrative changes and modifications to low-risk device registrations may qualify for immediate resolution.
  • Label changes: Updated Spanish labeling must be submitted when a modification affects the approved label.
  • Transfer of registration rights: Changes in ownership or transfer of product rights require the applicable notification or modification procedure.

The applicable procedure, supporting documents and response time depend on the nature of the change.

Conclusion

Successful IVD market entry in Mexico depends on seven elements. These are accurate classification, the correct authorization, a complete technical dossier, adequate analytical and clinical performance evidence, Spanish labeling that matches the approval, clear local responsibilities and continued post-market compliance.

Before submitting to COFEPRIS, manufacturers should review their classification, their documents and their Mexican applicant structure. Artixio provides medical device regulatory services in Mexico to help manufacturers navigate IVD classification, COFEPRIS registration, technical documentation, and market-entry compliance. For expert regulatory support, contact us at info@artixio.com.

FAQs

Q1. How are IVD products classified in Mexico?
IVDs fall into Class I, II or III. The class follows the intended purpose, the analyte and condition, the specimen, the user and the consequence of a wrong result.

Q2. Does every IVD product require COFEPRIS sanitary registration?
Most do. The exceptions are low-risk devices on the no-registration list and products that are not medical devices, both defined in the July 7, 2025 Acuerdo. Low-risk devices on the other list still need registration.

Q3. What documents are required for IVD registration in Mexico?
The application needs form FF-COFEPRIS-21, proof of payment, a Spanish draft label and instructions, and technical and scientific information per the COFEPRIS Guide. Foreign-made devices also need a certificate of free sale, GMP evidence and, in some cases, a manufacturer representation letter. It also needs proof of legal representation. The exact list depends on class and procedure.

Q4. Can a foreign manufacturer register an IVD product in Mexico?
Yes, through a Mexican applicant with a legal representative in Mexico. A Mexican registration holder should be planned for unless COFEPRIS confirms that a foreign company can hold the registration directly.